PMElogo.jpg (16797 bytes)
                                                                     Precision Measurement Engineering, Inc.


Discussing Tips & Techniques

PERSONAL DANGER
FLOOD
SHIPPING
SENSOR GUARD
CLOTHING
ROCKS
RELEASE CABLE
DAMAGED SENSORS
TESTS BEFORE A FIELD TRIP
USEFUL ITEMS (NOT SUPPLIED BY PME)

 

PERSONAL DANGER

If the SCAMP becomes substantially filled with water at depth, then it can return to the surface containing high pressure. If you have reason to believe that the SCAMP has flooded, such as failure to communicate or bubbling at any seal, then treat the SCAMP with extreme caution. Internal pressure may be released by carefully loosening the pressure test screw and allowing several minutes to pass. This can be done from the side using a screwdriver. Do not allow any part of your body to come into alignment with either endcap since these may unexpectedly be explosively expelled by internal pressure.


FLOOD

Should a SCAMP seal fail and allow the SCAMP to become flooded with water, open it, immediately drain the water, and remove all batteries. Remove the electronics covers and completely flush the covers and all circuits with large quantities of distilled water. Leave the SCAMP open so that it will dry. Allow it to dry completely, then close it and return it to PME for inspection.


SHIPPING

Save the cardboard shipping box  (12" x 12" x 36") that contains the SCAMP. This may be re-used to ship SCAMP from point to point or to return the SCAMP to PME.  Do not ship SCAMP in any other container.   If the cardboard box becomes unusable, then obtain another either from PME or some other convenient source.

When re-packing the SCAMP in the shipping box be sure to use all four pieces of foam. The two pieces with the hole in the middle of the SCAMP and the other two at either end of the box. This will help keep the SCAMP centered in the box.  Fill box with popcorn.

The grey flip pack will contain all the SCAMP elements except the SCAMP itself.   It takes a little arranging, but everything will fit. The tent stand PVC pipes lay on the bottom at one end. Place the cable assembly upright at the other end. Next, the drag plate unit fits well between the PVC pipes and the cable assembly. The guard can fit in this area too, but might need to be nested somewhat with the PVC pipes inserted into its interior. Make sure that it is positioned so that the supports do not become bent. The tent folds flat and can be placed on top.

SENSOR GUARD

When storing or shipping be sure that the sensor guard is securely packaged so that it cannot be bent. No disassembly is required.

After screwing the sensor guard onto the SCAMP, use a small cable tie or bit of dental floss to tie one of the guard's legs to one of the sensor seal nuts bases. This will prevent the unscrewing of the guard during a SCAMP deployment. Should the guard come unscrewed the retrieval cable can slip loose and the SCAMP will be lost. The sensor guards screw on tightly, but this extra precaution is low-cost insurance.


CLOTHING

You will find it difficult at best to see the laptop display in direct sunlight. Experience has shown that much of the glare on the laptop screen can come from light reflected from the operator's clothing. Wear something dark.

ROCKS

Tying the release cable to the rock is often a problem. Some suggestions are: cable tie the rock first, then tie to the cable tie, use a plastic zip lock bag to contain the rock, or use several rubber bands. Additional suggestions will be greatly appreciated.

RELEASE CABLE

Make these in advance of a field trip. They are only used for upwards profiling. There is a tool provided for creating the proper size loops to fit on the release screw.

Locate enough expendable weights (rocks) for the profiles you intend to make. Make wire loops from the wire provided by threading the wire through the crimp to make a loop, then adjusting the loop size using the tool provided. Crimp the collar in place with pliers. Connect the other ends of the wires to the expendable weights so that the weights will hang below the SCAMP by at least 12 inches (30cm).

scamp40.jpg (143811 bytes)


When you run out of  release cable and crimps you can re-order these from PME, purchase similar materials at your local fishing shop, or make release cables from 50# test fishing line. (For fishing line tie a slip-knot so that it tightens due to the weight of the rock. These aren't optimal since they do tend to slip off  if the rock tension is lost.)

PME part numbers for additional crimps and release cable are:

4580 - crimp, each  (please order in groups of 100 crimps)
4581 - release cable, roll of 100 feet

DAMAGED SENSORS

Damaged conductivity sensors should be replaced, or at least have dummy sensor plugs installed in their place.  Damage to Fast Conductivity sensors can admit water into the SCAMP by allowing it to travel down the Fast C glass tubes and along the connecting wires.  The conductivity circuits operate the sensor using an AC signal. In some cases, if the sensor is damaged or not installed, the circuit can enter an oscillating mode. This doesn't damage the circuit, but can induce a large amount of electrical noise in the other sensor channels in the SCAMP.  Dummy sensors must be electrically connected in place of damaged conductivity sensors.

Damaged temperature sensors need not be replaced since there are no likely leak paths.  Temperature circuits will not oscillate, but if the sensor is seriously damaged they can put out large voltages that will affect the other channels.  Damaged temperature sensors must be at least electrically disconnected and dummy sensors connected in their place.

Damaged oxygen sensors should be replaced because of leak possibilities along the wires within the sensor.  There is no dummy sensor required.  Simply disconnect the oxygen sensor.

See Replacing a Sensor section for more detailed information concerning sensor replacement and Installing a Dummy Sensor for dummy sensor information..

 

TESTING BEFORE A FIELD TRIP

Many features of the SCAMP should be tested prior to using it for data collection. These tests should be accomplished several weeks prior to the expected use to allow time for repairs and re-calibrations should these become necessary. Some of the tests should also be done on a daily basis or at other intervals. Many of these assume that the SCAMP is connected to the host computer and that the SCAMP Control Dialog is running.

Analog test. This test should be done in advance of a field trip.  Click the Display button in the Analog Test section of the Test Tab. See the Analog Test section for the proper test result.

Battery test. This test should be done in advance of a field trip and frequently during use. Simply review the System Tab battery indications.

Temperature sensor test. This test should be done in advance of a field trip and on a daily basis. If not already done, fill the sensor cover with tap water at ambient temperature or at the highest limit of SCAMP temperature channel calibration.  Rock the SCAMP back and forth to gently mix the water in the sensor cover. Make sure that it is completely mixed. Stop rocking the SCAMP and wait a moment. Record a brief profile with SCAMP.  Pour out the water in the sensor cover and replace it with ice water, or water cooled to the lowest limit of SCAMP temperature channel calibration. Be sure that no ice enters the cover as it can damage the fast sensors. Record another brief profile. Upload both profiles and review these using the Matlab software. Check that all temperature sensors agree within 0.050 deg C. If larger differences are observed at either temperature, then contact PME.

Conductivity sensor test. This test should be done in advance of a field trip and on a daily basis. Fill the sensor cover with salty water of conductivity near the upper limit of SCAMP conductivity channel calibration.  Rock the SCAMP back and forth to gently mix the water in the sensor cover. Make sure that it is completely mixed. Stop rocking the SCAMP and wait a moment.  Record a brief profile with SCAMP. Upload this and review using the Matlab software. The difference between Accurate and Fast Conductivity should be less than a few percent. If it is not, then carefully inspect the Fast Conductivity with a magnifier for fouling of the electrodes or cracks in the glass near the base or near the electrodes. Cracked sensors must be replaced. Fouled sensors can be cleaned (process being developed now). Re-calibrate the Fast Conductivity sensor using the Accurate Conductivity as a reference.

Fluorometer test (for SCAMPs with fluorometers installed). Record a brief profile with SCAMP while inserting and removing a blade of grass into the fluorometer measurement tube. Upload the profile and review using the Matlab software. The fluorometer should show a low voltage in the 0.1 to 0.5 Volt range with no grass in the tube and show a high voltage with grass inserted. The actual amplitudes will depend upon the sensitivity adjustment of the fluorometer.

Seal leakage test. This test should be done in advance of a field trip and after changing SCAMP batteries. For more information see the Testing for Leaks section.

If the above tests are all completed successfully, then nearly all of SCAMP's features will be verified.

USEFUL ITEMS (NOT SUPPLIED BY PME)

scissors for cutting lead ballast weights
duct tape for connecting the weights to the SCAMP
screwdrivers small standard and phillips
heavy dark cloth used to view the laptop computer; cover the computer tent and yourself to block the sun
pliers for squeezing the weight release line collars and operating the sensor seal nuts
voltmeter for checking battery voltages
cable ties for securing the guard against unscrewing
laptop computer power use a large capacity battery with an AC invertor so that the laptop can be powered by using its AC plug-in power supply
spare batteries for use in the SCAMP and especially the laptop
laptop serial port An extender cable is required for the laptop serial port since the SCAMP RS485-RS232 adaptor is too long to plug directly into the port. Allows the computer to easily fit into the tent.
squirt bottle fill with DI water to clean the sensor shafts and pressure transducer
float If the SCAMP becomes tangled on the bottom, then it may be necessary to leave it to obtain diving gear. Attach the float to the retrieval line to act as a marker.
o-ring grease for lubricating the various o-rings
spare sensors for replacing damaged sensors
panavise use two for holding the SCAMP to replace sensors or service the instrument