Discussing SCAMP Maintenance
PRESSURE TRANSDUCER
The pressure transducer is located within the release
endcap and receives external pressure through the 1/4" diameter hole drilled
approximately in the center of this cap. The pressure transducer should be carefully
flushed after deployment in salt water.
- Cut the nozzle of a squeeze bottle so that it will only
slightly penetrate the pressure transducer port. Be sure that the nozzle will not
penetrate so deeply that it touches the pressure transducer diaphragm. This can damage the
transducer.
- Gently squeeze distilled water into the pressure
transducer port. This water will be expelled from the pressure transducer cavity through
the small hole provided for this purpose drilled next to the port. Flush thoroughly.