Calibration
Each miniDOT is calibrated separately in PME's calibration chamber before being shipped to the customer. We calibrate at 12 oxygen concentrations each at 8 temperatures for a total of 96 calibration points. However we do not as yet use the sodium sulfite zero point in our calibration procedure. When miniDOT displays an oxygen measurement less than about 1 mg/l it is computing the value based upon an extrapolation of higher concentration oxygen measurement. These calibration extrapolations are subject to more error than the interpolations used at higher concentrations. miniDOTs read a slight negative value when exposed to zero oxygen. We account for this by specifying a overall absolute error as well as a fractional error for miniDOT performance.
Please send calibration questions to info@pme.com.
|