Sensor Specifications

 

Fast Conductivity

Measurement equation
Vo = G* C-5
8 hour calibration stability
Better than 1% of C reading
Time response
-3db at approximately 800 Hz
Spatial response
-3db at approximately 4 cycles/cm
Noise
<1 mv RMS 10 Hz to 1 KHz
Output voltage range
+/-5 volts
Linear conductivity range
0.5 to 800 mS/cm
Calibration
performed by user

Fast Temperature

Measurement Equation
Vo = G exp (A + B/T) - 5
8 hour calibration stability
better than 0.05 degrees C
Time response
0.007 second time constant
Noise
< 1 mv RMS 10 Hz to 200 Hz
Output voltage range
+/- 5 volts
Temperature range
0 to 40 degrees C
Calibration
performed by PME (0 to 30 degrees C)

Request MSCTI Quote

 

 
 

Precision Measurement Engineering, Inc.
2792 Loker Ave. West Ste. 105
Carlsbad, CA 92010
(888) 841-7464